Renma® - PTA balloon dilatation catheter

0.035" RX (rapid exchange) design

  • Rapid exchange facilitates fast and convenient device exchange, saves radiation dose and enables the use of short guidewires
  • Perfect match with the self-expanding peripheral stent Misago, which also features RX design

Semi-compliant 6-folded balloon with SiLX² coating

  • Low crossing profile and great re-folding capability for large diameters (Ø 7-10 mm)
  • SiLX² coating for smooth trackability and easy retrieval after multiple inflations

Dedicated shaft design

  • Combines the pushability of a stainless steel hypotube shaft with the kink resistance of a polymer shaft
  • For use in even cross-over procedures in sharp, angulated bifurcations
  • Maintains very fast deflation times
Read more Read less
Catheter designRapid exchange, coaxial polymer/hypotube shaft
Guidewire compatibility0.035" (0.89 mm)
Usable shaft length135 cm
Balloon materialPolyamide (Nylon)
Balloon folding3 folded (Ø 4 – 6 mm), 6 folded (Ø 7 – 10 mm)
CoatingSiLX² coating
Shaft diameter​Proximal Part: 2.6 Fr (0.88mm),

Distal Part: 4.0 – 7.0 mm 5.0 Fr (1.65 mm), 8.0 – 10.0 mm 5.2 Fr (1.73 mm)

Introducer compatibility6 Fr (2.01 mm)
Nominal pressure (NP)6 atm / 608 kPa

Balloon

Diameter

(mm) ​ ​
Item reference ​ ​Rated Burst Pressure (RBP) ​
Balloon length (mm) ​ ​ ​ ​Balloon length (mm) ​
2040608010020, 40 and 6080 and 100
4.0BD-C40201DNBD-C40401DNBD-C40601DNBD-C40801DNBD-C40A01DN16 atm / 1621 kPa15 atm / 1520 kPa
5.0BD-C50201DNBD-C50401DNBD-C50601DNBD-C50801DNBD-C50A01DN14 atm / 1419 kPa13 atm / 1317 kPa
6.0BD-C60201DNBD-C60401DNBD-C60601DNBD-C60801DNBD-C60A01DN13 atm / 1317 kPa12 atm / 1216 kPa
7.0BD-C70201DNBD-C70401DNBD-C70601DNBD-C70801DNBD-C70A01DN12 atm / 1216 kPa12 atm / 1216 kPa
8.0BD-C80201DNBD-C80401DNBD-C80601DN12 atm / 1216 kPa
9.0BD-C90201DNBD-C90401DNBD-C90601DN11 atm / 1115 kPa
10.0BD-CA0201DNBD-CA0401DNBD-CA0601DN

11 atm / 1115 kPa

Please quote above item reference codes when placing an order.

​This Product is no longer available for sales.

​This Product is no longer available for sales.

You will find here electronic copies of Instructions for Use for selected Terumo products to which European Regulation No. 207/2012 has been applied.

Below you can download the actual Instruction Manual for your product, up to 15 years at least after purchase.

All documents are available to download as PDF files. To read them, you must have Adobe Reader* installed on your computer. If necessary, click here to download Adobe Reader for free.

Please contact Terumo to receive a printed copy of one of the listed e-IFU's within 7 days.

The version of the Instruction Manual that you should refer to depends on what product you received. Please check the Serial Number or lot number of your product to determine the version that goes along with it.

Title Revision number Item reference Revision date Serial number
{{download.Title}} {{download.RevisionNr}} {{download.ItemRef}} {{download.RevisionDate}} {{download.SerialNr}}

 

 

 

 

 

 

 

 

An Error Occured In Hayabusa SeeAlso: Object reference not set to an instance of an object.

Did you not find the product you were looking for?