  • ​Provides impressive high-pressure resistance up to 22 atm
  • Optimizes (re)crossability and deliverability1
  • Operates accurately in pre- and post-dilatation
Read more Read less

Shaft diameter

1.9 Fr / 0.63 mm Proximal
2.5 Fr / 0.84 mm Mid
2.6 Fr / 0.86 mm Distal

Usable length

145 cm

Balloon markers

2

Nominal pressure

12 atm / 1216 kPa

Rated burst pressure

22 atm / 2229 kPa (2.0 mm to 4.0 mm)
20 atm / 2026 kPa (4.5 and 5.0 mm)

Guidewire compatibility

Maximum diameter 0.014" / 0.36 mm

Lesion entry profile

0.43 mm

Coating

Hydrophilic

Ordering Information

Balloon
diameter
(mm) ​
Balloon length (mm) / Item references  ​ ​ ​ ​ ​ ​
681215202530
2.00DC-RM2006HSWDC-RM2008HSWDC-RM2012HHWDC-RM2015HHWDC-RM2020HHW
2.25DC-RM2206HSWDC-RM2208HSWDC-RM2212HHWDC-RM2215HHWDC-RM2220HHW
2.50DC-RM2506HSWDC-RM2508HSWDC-RM2512HHWDC-RM2515HHWDC-RM2520HHWDC-RM2525HHWDC-RM2530HHW
2.75DC-RM2706HSWDC-RM2708HSWDC-RM2712HHWDC-RM2715HHWDC-RM2720HHW
3.00DC-RM3006HSWDC-RM3008HSWDC-RM3012HHWDC-RM3015HHWDC-RM3020HHWDC-RM3025HHWDC-RM3030HHW
3.25DC-RM3206HSWDC-RM3208HSWDC-RM3212HHWDC-RM3215HHWDC-RM3220HHW
3.50DC-RM3506HSWDC-RM3508HSWDC-RM3512HHWDC-RM3515HHWDC-RM3520HHWDC-RM3525HHWDC-RM3530HHW
3.75DC-RM3706HSWDC-RM3708HSWDC-RM3712HHWDC-RM3715HHWDC-RM3720HHW
4.00DC-RM4006HSWDC-RM4008HSWDC-RM4012HHWDC-RM4015HHWDC-RM4020HHW
4.50DC-RM4506HSWDC-RM4508HSWDC-RM4512HSWDC-RM4515HSW
5.00DC-RM5006HSWDC-RM5008HSWDC-RM5012HSWDC-RM5015HSW

Please quote above item reference codes when placing an order.

Balloon Compliance

Balloon
diameter
(mm) ​
Pressure (atm) ​ ​ ​ ​ ​ ​ ​ ​ ​ ​ ​ ​ ​ ​ ​ ​ ​ ​ ​ ​ ​ ​
345678910111213141516171819202122232425
2.001.581.651.721.791.841.881.921.951.982.00*2.012.032.042.052.062.082.092.102.122.13**2.142.152.17
2.251.962.012.062.112.142.162.192.212.232.25*2.262.282.292.312.322.342.352.362.382.39**2.412.422.44
2.502.192.242.302.362.382.412.442.462.482.50*2.522.532.552.562.582.602.612.632.642.66**2.682.692.71
2.752.432.482.532.592.622.652.672.702.732.75*2.772.792.802.822.842.862.872.892.912.93**2.942.962.98
3.002.652.712.762.822.852.892.922.952.973.00*3.023.043.063.083.103.113.133.153.173.19**3.213.233.25
3.252.892.953.013.073.103.143.173.203.223.25*3.273.293.313.333.353.373.393.423.443.46**3.473.483.51
3.503.133.193.263.323.353.393.423.453.473.50*3.523.543.573.593.613.633.653.683.703.72**3.743.743.76
3.753.373.443.503.563.603.633.663.693.723.75*3.773.803.823.843.863.883.903.923.943.96**3.983.994.01
4.003.583.653.723.793.833.873.903.943.974.00*4.034.054.074.094.114.134.154.174.194.21**4.234.264.28
4.504.024.104.194.284.324.374.404.444.474.50*4.534.564.594.634.664.684.714.75**4.784.814.844.874.91
5.004.524.594.674.754.804.854.894.934.965.00*5.035.075.105.135.165.195.225.26**5.295.325.355.385.41

* Nominal pressure 
** Rated burst pressure

 

 

 

 

 

 

 

 